Description

LipoClear is a product that can allows for measurement of most chemistry tests in a lipemic specimen.


 

Analytes that cannot be done after LipoClear:

(1) lipid studies

(2) coagulation tests

(3) monoclonal protein

 

LipoClear comes as a liquid. All results must be multiplied to compensate for dilution.

 

actual analytical concentration =

= 1.2 * (measured concentration)

 

Exceptions:

(1) Total protein should be increased a further 0.6 g/dL (range 0.4 to 0.8) due to the loss of lipoproteins.

(2) Each electrolyte must be adjusted using a factor based on the electrolyte and the analytical system.

(3) Phosphorus may be increased spuriously by 10% in some analytical systems. The correction factor for the measured concentration in this case should be 1.09 (with 1.09 * 1.10 = 1.20).

 