Description

Ruiz et al listed grading criteria for a paravalvular leak with a prosthetic aortic valve. The authors are from multiple institutions in North America and Europe.


Patient selection: paravalvular leak for a prosthetic aortic valve

 

Parameters:

(1) sewing ring motion

(2) jet features

(3) percent left ventricular outflow tract (LVOT) diameter

(4) circumferential extent

(5) left ventricular size

(6) vena contracta width in mm

(7) spectral Doppler

(8) pressure half time (PHT) in milliseconds

(9) diastolic flow reversal

(10) right ventricular volume

(11) regurgitant fraction in percent

(12) effect regurgitant orifice area (EROA) in square cm

 

Parameter

Mild Leak

Severe Leak

sewing ring motion

normal

usually abnormal

jet features

narrow jet width, infrequent multiple, no proximal flow convergence

wide jet width, frequently multiple, proximal flow convergence

percent LVOT diameter

< 30%

>=60%

circumferential extent

<10%

>=30%

left ventricular size

normal

moderately to severely dilated

vena contracta width

< 4 mm

>= 6 mm

spectral Doppler

incomplete or faint

dense

PHT in milliseconds

> 500 ms

< 200 ms

diastolic flow reversal

absent or brief

holodiastolic

right ventricular volume

< 30 mL

>= 60 mL

regurgitant fraction

< 30%

>= 50%

EROA in square cm

< 0.1 square cm

>= 0.3 square cm

 