Description

The dead volume of a chromatographic column can be calculated from the column dimensions. If the flow rate is known, then the dead time can be estimated.


 

Parameters:

(1) length of the column in cm

(2) inner diameter of the column in cm

(3) flow rate through the column

 

For the column:

 

dead volume in mL =

= 0.5 * ((diameter in cm)^2) * (column length in cm)

 

If the flow rate is known, then:

 

dead time in minutes =

= (dead volume in mL) / (flow rate in mL/min)

 

The retention factor for a compound can be calculated as:

 

retention factor =

= ((retention time for the compound) - (dead time)) / (dead time)

 

If the retention factor is between 1 and 20 then the chromotographic system should give acceptable performance.

 