The size of gel chromatography column can be determined by the separation needs.
Goal |
Column Size in mL |
separation to 20 g |
100 * (separation sample size ini grams) |
separation from 21 to 250 g |
80 * (separation sample size in grams) |
separation from 250 to 300 g |
100 * (separation sample size ini grams) |
filtration to 250 grams |
40 * (filtration sample size in grams) |
filtration from 250 to 300 g |
66 * (filtration sample size in grams |
upper limit of fraction size in mL =
= (column size in mL) / 20
For a column above 2,000 mL the lower limit to the fraction size may be 25 to 33% less.
A more complex equation for columns 4,000 to 20,00 mL
fraction size in mL =
= (-0.01074 * ((column size) / 100)^2) + (5.969 * (column size) / 100) - 13.63