Description

The size of gel chromatography column can be determined by the separation needs.


Goal

Column Size in mL

separation to 20 g

100 * (separation sample size ini grams)

separation from 21 to 250 g

80 * (separation sample size in grams)

separation from 250 to 300 g

100 * (separation sample size ini grams)

filtration to 250 grams

40 * (filtration sample size in grams)

filtration from 250 to 300 g

66 * (filtration sample size in grams

 

upper limit of fraction size in mL =

= (column size in mL) / 20

 

For a column above 2,000 mL the lower limit to the fraction size may be 25 to 33% less.

 

A more complex equation for columns 4,000 to 20,00 mL

 

fraction size in mL =

= (-0.01074 * ((column size) / 100)^2) + (5.969 * (column size) / 100) - 13.63

 